Isolation Transformer Model EIT-2000
|
Designed and built to comply with the requirements of Medical Equipment Specification, BS EN 60601-1
|
General Specification
|
 |
2000VA Medical Toroidal Isolation Transformer |
 |
 |
IEC inlet fused in each line |
 |
 |
Medical filter on input |
 |
 |
IEC Outlets 6 way |
 |
 |
Thermal Circuit Breaker illuminated |
 |
 |
Inrush current limiting |
 |
 |
Fused outputs |
 |
 |
Metal case 260mm W x 345mm D x 130mm H |
 |
 |
Carrying handles |
 |
Transformer Electrical Specification
|
 |
Primary : 0-230V 50/60Hz |
 |
 |
Secondary : 0-230V @2000VA maximum |
 |
 |
Regulation : <4% |
 |
 |
Temperature Rise : <40°C |
 |
 |
Leakage : Primary to secondary <100µA @264V 50Hz |
 |
 |
Leakage : Primary to screen <500µA @ 264V 50Hz |
 |
 |
Flash Test: Primary to screen @2kV rms |
 |
 |
Flash Test : Primary to secondary @ 5kV rms |
 |
 |
Overall Specification
|
 |
Input Voltage : 220V to 240V nominal 50/60Hz |
 |
 |
Input Power : 2100W |
 |
 |
Input Filter : Medical low leakage <5µA rating 10A
|
 |
 |
Input Fuses : Minature fuse 20mm x 5mm 10A anti surge type
|
 |
 |
Output Rating : 220 to 240V @ 2000VA maximum |
 |
 |
Output Fuses : Miniature fuse 20mm x 5mm 10A anti surge type
|
 |
 |
Inrush Current : Thermistor limited 24A @264V
|
 |
 |
Power On Indication : Incorporated in circuit breaker |
 |
 |
Output Indicator : Neon green inserted between fuses and output sockets |
 |

|